| Model Number |
10603 |
| Medical Device Problem Codes |
Failure to Advance (2524); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that stent damage occurred.The target lesion was located in a coronary artery.A 3.00 x 20mm synergy drug-eluting stent was advanced for treatment.However, it was noted that the device was unable to cross the lesion and the struts were damaged.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.The target lesion was located in a coronary artery.A 3.00 x 20mm synergy drug-eluting stent was advanced for treatment.However, it was noted that the device was unable to cross the lesion and the struts were damaged.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.It was further reported that the target lesion was 90% stenosed, severely tortuous and moderately calcified.The lesion was pre-dilated with a 2.0x20 emerge balloon catheter.
|
| |
|