| Model Number |
7210569F |
| Medical Device Problem Code |
Decrease in Pressure (1490)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the spider was not holding pressure during an inspection.There was a smith and nephew backup device available.No other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product.No issues were noted.A functional evaluation was performed.The device failed the weight test.The piston migration was at 2.3 inches.The reported failure was confirmed and the root cause has been associated with a seal failure.A review of device records shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
| |
|