| Model Number |
H7493941824250 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 2.50 x 24mm synergy xd drug-eluting stent was advanced for treatment.However, during the procedure, the proximal end of the stent appeared to be frayed.The device was completely removed from the patient's body.The procedure was completed with another of same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: a synergy xd mr us 2.50 x 24mm stent delivery system was returned for analysis.A visual examination of the stent found evidence of damage with struts on the proximal section in a lifted state.The undamaged crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 2.50 x 24mm synergy xd drug-eluting stent was advanced for treatment.However, during the procedure, the proximal end of the stent appeared to be frayed.The device was completely removed from the patient's body.The procedure was completed with another of same device.There were no patient complications reported.
|
| |
|
Search Alerts/Recalls
|