| Model Number |
0279250101 |
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip broke during the procedure.The piece was retrieved.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip broke during the procedure.The piece was retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the ceramic on the tip was broken off.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be excessive force used when inserting removing the probe, a probe inserted into an obstructed passageway, or any forceful impact to the tip of the probe with rigid objects.The probe is used as a tool for a mechanical displacement of tissue or bone, or to pry or scrub.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Search Alerts/Recalls
|