| Model Number |
1217-01-054 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The thread of the cup seems to be oversized.The apex hole eliminator could be rotated beyond the cup dome.There was no significant surgery delay.No implant had to be removed.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated that the affected side was the right side.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h8.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot = > 1) quantity manufactured: (b)(4).2) date of manufacture: 25-apr-2021.3) any anomalies or deviations identified in dhr: none.4) expiry date: 31-mar-2031.5) ifu reference: 090200701.
|
| |
|
Search Alerts/Recalls
|