MAUDE Adverse Event Report: COOK ENDOSCOPY ACUSNARE POLYPECTOMY SNARE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY ACUSNARE POLYPECTOMY SNARE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number G22629
Medical Device Problem Code Failure to Conduct (1114)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter; occupation: non-healthcare professional.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The customer provided two pictures.One picture is of the pouch and the lot number and rpn match what's in the report.The second picture is of the device in the pouch, the picture is inconclusive for "snare cannot cut the polyps." the device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use contain the following information to assist with proper set-up and use of the device: "before using this device, follow recommendations provided by electrosurgical unit manufacturer to ensure patient safety through proper placement and utilization of patient return electrode.Ensure a proper path from patient return electrode to electrosurgical unit is maintained throughout the procedure." "inspect active cord.Cord must be free of kinks, bends, breaks and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." "with electrosurgical unit off, prepare equipment.Securely connect active cord to device handle and electrosurgical unit.Active cord fittings should fit snugly into both device handle and electrosurgical unit.Following instructions from electrosurgical unit manufacturer, position patient return electrode and connect it to electrosurgical unit." "following electrosurgical unit manufacturer's instructions for settings, verify desired settings and activate electrosurgical unit.Note: maximum rated input voltage for this device is 2kvp-p for cut mode and 5 kvp-p for coagulation mode." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During a polypectomy, the physician used an acusnare polypectomy snare.It was reported that the user opened the package and advanced the device to desired position and plugged the device with power, but found when electrosurgical generator is activated, no electrosurgical current was present at the snare head and cannot remove the polyps.The user then changed to another same device to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Continued: section e.Initial reporter; occupation: non-healthcare professional.Investigation evaluation: the customer provided two pictures.One picture is of the pouch and the lot number and rpn match what's in the report.The second picture is of the device in the pouch, the picture is inconclusive for "snare cannot cut the polyps." our laboratory evaluation of the product said to be involved could not confirm the report as it was described.The device was returned with the snare fully retracted into the tubing.The snare was advanced and retracted using the handle with no issues.During a visual inspection of the device, there was no damage noted.The continuity from the electrical pin to the snare head was tested with an ohm meter and passed.An additional functional test was performed by attaching the active cord to the electrical pin.The active cord connected easily and remained securely connected to the device.The device was connected to a valley lab generator and power was applied.The snare cut simulated tissue as expected.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for the reported observation could not be determined because the product said to be involved functioned as intended.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.The instructions for use contain the following information to assist with proper set-up and use of the device: "before using this device, follow recommendations provided by electrosurgical unit manufacturer to ensure patient safety through proper placement and utilization of patient return electrode.Ensure a proper path from patient return electrode to electrosurgical unit is maintained throughout the procedure." "inspect active cord.Cord must be free of kinks, bends, breaks and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." "with electrosurgical unit off, prepare equipment.Securely connect active cord to device handle and electrosurgical unit.Active cord fittings should fit snugly into both device handle and electrosurgical unit.Following instructions from electrosurgical unit manufacturer, position patient return electrode and connect it to electrosurgical unit." "following electrosurgical unit manufacturer's instructions for settings, verify desired settings and activate electrosurgical unit.Note: maximum rated input voltage for this device is 2kvp-p for cut mode and 5 kvp-p for coagulation mode." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Additional Manufacturer Narrative
A correction was made to the manufacturing and expiration dates.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUSNARE POLYPECTOMY SNARE
Common Device Name
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key12041424
Report Number1037905-2021-00261
Device Sequence Number3925879
Product Code KNS
UDI-Device Identifier00827002226296
UDI-Public(01)00827002226296(17)230928(10)W4389607
Combination Product (Y/N)N
PMA/510(K) Number
K851958
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2023
Device Model NumberG22629
Device Catalogue NumberAS-1-S
Device Lot NumberW4389607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2021
Initial Date Received by Manufacturer 05/28/2021
Supplement Date Received by Manufacturer06/21/2021
03/23/2022
Initial Report FDA Received Date06/22/2021
Supplement Report FDA Received Date07/16/2021
04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENDOSCOPE, UNKNOWN MAKE OR MODEL.; ERBE ELECTROSURGICAL UNIT.; OLYMPUS ENDOSCOPE.
Patient Age38 YR
Patient SexFemale
Patient Weight59 KG
-
-