• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NELLCOR PURITAN BENNETT IRELAND INVOS; OXIMETER, TISSUE SATURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NELLCOR PURITAN BENNETT IRELAND INVOS; OXIMETER, TISSUE SATURATION Back to Search Results
Model Number 5100C-EU
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during an open heart surgery, while the patient was on the heart-lung bypass machine, the hcp's noticed a bilateral decline in rso2 reading, with no specific reason.Actions were taken like deeper induction, higher flow rate extension and a mild improvement was seen.Hcp decided to switch to a different rso2 device, showed a higher base line value 61-56 (casmed) vs 42-36 (invos).There was no patient injury.
 
Additional Manufacturer Narrative
H3: evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted there was a defective component.It was reported that the monitor provided lower oximetry readings than expected.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVOS
Common Device Name
OXIMETER, TISSUE SATURATION
Manufacturer (Section D)
NELLCOR PURITAN BENNETT IRELAND
micheal collins rd mervue
galway
EI 
MDR Report Key12041572
Report Number8020893-2021-00092
Device Sequence Number10987359
Product Code MUD
UDI-Device Identifier10884521211179
UDI-Public10884521211179
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K091224
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5100C-EU
Device Catalogue Number5100C-EU
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/31/2021
Supplement Date Received by Manufacturer08/05/2021
Initial Report FDA Received Date06/22/2021
Supplement Report FDA Received Date08/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age14 YR
-
-