• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. PICO 7 DAY SYSTEM MULTISITE LG; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. PICO 7 DAY SYSTEM MULTISITE LG; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801804
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/04/2021
Event Type  malfunction  
Event Description
It was reported that a pico 7 day system multisite lg displayed air leak alarm.There was no apparent cracks in the device, nor tears in the tubing.It is believed the issue was provoked by incorrect dressing adhesion.Application of adhesive tape around dressing edges was recommended.Further information is not available.
 
Manufacturer Narrative
The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.While a leak in the npwt system (pump/tubing/dressing) may lead to a loss of negative pressure, it will not directly cause or contribute to serious injury or death even in a clinical setting as the pico dressing can revert to managing the wound as a standard multilayer dressing.No risk to patient safety is anticipated.This event is considered not reportable pursuant to 21 cfr part 803.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICO 7 DAY SYSTEM MULTISITE LG
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12042910
MDR Text Key257519567
Report Number8043484-2021-01545
Device Sequence Number1
Product Code OMP
UDI-Device Identifier05000223490461
UDI-Public5000223490461
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66801804
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/08/2021
Initial Date FDA Received06/22/2021
Supplement Dates Manufacturer Received09/16/2021
Supplement Dates FDA Received09/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-