| Model Number |
AR310 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Root cause: potential for alteration of staining in this case was due to improper operation of the heater on the slide carousel.The problem was solved by field service engineer with repair of the part.Following the repair, the instrument was fully operational within specifications, without errors and available for the user.Failure mode description: following a heater malfunction or if the part stops working; the resulting failure modes described could occur:, heater cable malfunction and temperature misalignment.Failure mode in all scenarios has the potential to cause a staining alteration.
|
| |
|
Event or Problem Description
|
|
Customer complaint record reported the event as follows: heaters do not reach temperature.No direct or indirect patient harm or user harm have been reported.
|
| |
|
Search Alerts/Recalls
|