• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERJECT; BIOPSY NEEDLE KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERJECT; BIOPSY NEEDLE KIT Back to Search Results
Model Number M00518301
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an interject needle was opened on (b)(6) 2021.During preparation, the physician found hair contamination.It was reported that the device was not unpacked.There was no patient involvement in this case.
 
Additional Manufacturer Narrative
H6: medical device problem code a1802 captures the reportable event of packaging foreign matter.H10: investigation results the returned unpackaged optiflo hemostasis catheter was analyzed, a visual evaluation was performed and a hair was observed inside the package.No other issues were noted.Based on all available information and the condition of the returned device, problems traced to the failure to maintain or establish techniques for controlling and verifying the product specifications (including materials used) identified by the manufacturer himself.The investigation concluded the most probable cause is quality control deficiency.There is an investigation in place to address this issue.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that an interject needle was opened on (b)(6) 2021.During preparation, the physician found hair contamination.It was reported that the device was not unpacked.There was no patient involvement in this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERJECT
Common Device Name
BIOPSY NEEDLE KIT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key12044305
Report Number3005099803-2021-02781
Device Sequence Number10987370
Product Code FCG
UDI-Device Identifier08714729296508
UDI-Public08714729296508
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K171454
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/17/2024
Device Model NumberM00518301
Device Catalogue Number1830
Device Lot Number0026647311
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2021
Initial Date Received by Manufacturer 05/26/2021
Supplement Date Received by Manufacturer06/29/2021
Initial Report FDA Received Date06/22/2021
Supplement Report FDA Received Date07/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-