• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 550; CARIDOPULMONARY BYPASS ON-LINE BLOOD GAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 550; CARIDOPULMONARY BYPASS ON-LINE BLOOD GAS Back to Search Results
Model Number 550AVHCT
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported complaint was confirmed.The srt noticed during oven testing that on the last cycle of four hours +/- 10% (24 minutes) at 30 degrees celsius +/- 5 degrees celsius, the venous probe failed the test with a temperature reading of 6.6 degrees celsius, which is out of specification.The srt replaced the venous probe due to intermittent thermistor readings during the oven testing and the unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event or Problem Description
The service repair technician (srt) reported that during testing of the device at the service center, the blood parameter monitor (bpm) venous probe failed the oven test with an out of specification temperature reading.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 550
Common Device Name
CARIDOPULMONARY BYPASS ON-LINE BLOOD GAS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key12044402
Report Number1828100-2021-00215
Device Sequence Number4110439
Product Code DRY
UDI-Device Identifier00886799001783
UDI-Public(01)00886799001783(11)190316
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K182110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number550AVHCT
Device Catalogue Number550AVHCT
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/28/2021
Initial Report FDA Received Date06/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-