| Model Number |
550AVHCT |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported complaint was confirmed.The srt noticed during oven testing that on the last cycle of four hours +/- 10% (24 minutes) at 30 degrees celsius +/- 5 degrees celsius, the venous probe failed the test with a temperature reading of 6.6 degrees celsius, which is out of specification.The srt replaced the venous probe due to intermittent thermistor readings during the oven testing and the unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
The service repair technician (srt) reported that during testing of the device at the service center, the blood parameter monitor (bpm) venous probe failed the oven test with an out of specification temperature reading.There was no patient involvement.
|
| |
|