NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE PLATING SYSTEM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number LP4.5-PRF4B
Medical Device Problem Codes Corroded (1131); Material Discolored (1170)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been received and the evaluation is pending.A supplemental report will be provided once the evaluation is complete.
 
Event or Problem Description
Information was received that during implant removal procedure, it was discovered that the plate had visible discoloration and corrosion.At the time of removal, a low profile nail was inserted for extra stability.The desired length was achieved.
 
Additional Manufacturer Narrative
Device evaluation: visual inspection of the returned plate revealed the distraction rod of the plate had discoloration which confirmed the reported failure mode.Per the reported failure functional testing was not applicable.Device records review: a review of device history records for the returned unit confirmed that it met all the required quality inspections prior to release.
 
Event or Problem Description
No additional information was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE PLATING SYSTEM
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key12044921
Report Number3006179046-2021-00316
Device Sequence Number3821006
Product Code KTT
UDI-Device Identifier00887517021366
UDI-Public887517021366
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K192181
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLP4.5-PRF4B
Device Lot Number0072003AAA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/01/2021
Supplement Date Received by Manufacturer08/16/2021
Initial Report FDA Received Date06/22/2021
Supplement Report FDA Received Date08/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age8 YR
-
-