ENDOLOGIX AFX; VELA SUPRARENAL Back to Search Results
Model Number A34-34/C80-O20 V
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.
 
Event or Problem Description
The patient was initially implanted with an afx2 bifurcated stent graft, a suprarenal aortic extension, and an ovation ix extension to treat an abdominal aortic aneurysm (aaa).This initial procedure is outside the indications of use (off-label) due to the use of adjunctive devices not compatible with afx system per the ifu.Approximately one (1) year post initial procedure, during a routine follow up, a type 1a endoleak was identified.The patient is doing fine and is currently being monitored while an intervention is being discussed.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type 1a endoleak is unconfirmed.This is not consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy-relatedness of this complaint could not be determined with the medical records available for review.The final patient status was reported as doing fine and is currently being monitored while intervention is being discussed.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply.Corrections: g4: date of received by manufacturer.H6: result code ¿ remove 3233.H6: conclusion code ¿ remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
VELA SUPRARENAL
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key12045992
Report Number2031527-2021-00239
Device Sequence Number11816714
Product Code MIH
UDI-Device Identifier00818009013606
UDI-Public(01)00818009013606(17)220926
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/26/2022
Device Model NumberA34-34/C80-O20 V
Device Lot Number2290790-002
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/10/2021
Supplement Date Received by Manufacturer08/09/2021
Initial Report FDA Received Date06/22/2021
Supplement Report FDA Received Date08/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX2 BIFURCATED STENT GRAFT, LOT # 1737476-009; OVATION IX EXTENDER, LOT # FS030420-41
Outcome Attributed to Adverse Event Other;
Patient Age88 YR
-
-