• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS SA2.0 20 & PED.PRECHAMBER; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS SA2.0 20 & PED.PRECHAMBER; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX596T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 05/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: visual inspection: deposits in the prechamber were observed through the visual inspection.No significant deformations or damage were detected.Permeability test: a permeability test has shown that both valves are permeable.Some particles are flushed out during this test.Computer controlled test: to investigate the claim of over/under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves operated within the accepted tolerance in both positions.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Results: first, we performed a visual inspection of the progav 2.0 shunt system.In the prechamber were some deposits visible, but no significant deformations or damage of the valves were detected, during the visual inspection.Next, we tested the permeability, adjustability, and opening pressure of the valve, as well as the brake functionality and brake force.The valves operated as expected and met all specifications.Finally, we have dismantled the valves.Inside both valves we have found build-up of substances (likely protein).Based on our investigation, we are unable to substantiate the clinical claim of occlusion, under- and over-drianage.At the time of our investigation, the valves were shown to be permeable and operated within the specifications.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a progav 2.0 shuntsytsem (part # fx596t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the valve was believed to be operating in over/under drainage and/or clogged.The patient underwent a revision procedure on (b)(6) 2021.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6) height: 65 centimeters (cm) weight: (b)(6) gender: unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 SYS SA2.0 20 & PED.PRECHAMBER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
MDR Report Key12046630
Report Number3004721439-2021-00166
Device Sequence Number11809032
Product Code JXG
UDI-Device Identifier04041906503536
UDI-Public4041906503536
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K190174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model NumberFX596T
Device Catalogue NumberFX596T
Device Lot Number20048714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2021
Initial Report FDA Received Date06/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age2 MO
Patient Weight5
-
-