Investigation: visual inspection: deposits in the prechamber were observed through the visual inspection.No significant deformations or damage were detected.Permeability test: a permeability test has shown that both valves are permeable.Some particles are flushed out during this test.Computer controlled test: to investigate the claim of over/under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves operated within the accepted tolerance in both positions.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Results: first, we performed a visual inspection of the progav 2.0 shunt system.In the prechamber were some deposits visible, but no significant deformations or damage of the valves were detected, during the visual inspection.Next, we tested the permeability, adjustability, and opening pressure of the valve, as well as the brake functionality and brake force.The valves operated as expected and met all specifications.Finally, we have dismantled the valves.Inside both valves we have found build-up of substances (likely protein).Based on our investigation, we are unable to substantiate the clinical claim of occlusion, under- and over-drianage.At the time of our investigation, the valves were shown to be permeable and operated within the specifications.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|