| Model Number |
ZLB00 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age, weight, ethnicity: information asked/not provided.If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.An attempt to obtain additional information was conducted, however there is no additional information available.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the intraocular lens (iol) had a crack.The iol was cut out and removed from the eye during the same procedure.Another johnson & johnson lens (same model and diopter) was being used as the replacement lens, however while being implanted the patient's capsule broke.Through follow-up, it was confirmed there was no vitrectomy, incision enlargement, or sutures required in the patient's left eye.No additional information was provided to johnson & johnson surgical vision.This report is being submitted to capture the cracked iol.A separate report is being submitted for the capsule break event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: 8/25/2021.Section h3: device evaluated by manufacturer¿ yes.Device evaluation: the complaint lens was received in the original lens case.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptic and that the lens was received cut in half (only one half received), which is consistent with a lens that was handled during removal and replacement.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed and no product deficiency could be identified.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device was not returned at the manufacturing site; therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.The search revealed that no similar complaint for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|