• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1250 SYNCHRONY 2 PIN
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/21/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The clinic reported that in situ testing of the device shows fluctuating impedances.
 
Additional Manufacturer Narrative
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
 
Event or Problem Description
The clinic reported that in situ testing of the device shows fluctuating impedances.The user has been re-implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key12047843
Report Number9710014-2021-00460
Device Sequence Number3427241
Product Code MCM
UDI-Device Identifier09008737366780
UDI-Public(01)09008737366780
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/20/2023
Device Model NumberMI1250 SYNCHRONY 2 PIN
Device Catalogue Number36707
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/21/2021
Supplement Date Received by Manufacturer06/21/2021
Initial Report FDA Received Date06/23/2021
Supplement Report FDA Received Date10/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age3 YR
-
-