• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT 5F DUAL-LUMEN REVERSE TAPER PICC BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT 5F DUAL-LUMEN REVERSE TAPER PICC BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported that because an alarm to notify infusion pump occlusion sounded, the catheter and the insertion site were checked.Blood return from both lumens of the red and white were not confirmed, and leakage occurred around the insertion site after saline solution was infused.When the dressing was removed, the catheter was kinked and a crack was found on the catheter.Therefore, the device was removed from the patient¿s body.There was no reported patient injury.
 
Event or Problem Description
It was reported that because an alarm to notify infusion pump occlusion sounded, the catheter and the insertion site were checked.Blood return from both lumens of the red and white were not confirmed, and leakage occurred around the insertion site after saline solution was infused.When the dressing was removed, the catheter was kinked and a crack was found on the catheter.Therefore, the device was removed from the patient¿s body.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed and the cause appears to be use related.The product returned for evaluation was one 5fr d/l groshong catheter.The damage observed in the returned sample was characteristic of over-pressurization (burst) damage.This can occur through the use of syringes smaller than 10ml, by flushing against an occlusion, or due to excessive force applied during infusion procedures.The returned product sample was evaluated and a split was observed in the white-luered lumen, at the 49cm depth marking.This catheter damage was typical of a burst, and the characteristics observed which supported this type of failure included: 'c' shaped split.Tensile weakness at the fracture site (due to material ballooning prior to burst).Granular fracture surface texture (typical of tearing failure modes).The catheter material was visibly lighter in color at the fracture site.A partial occlusion was observed distal to the burst site, and this may have contributed to the failure.Forceful flushing against an occlusion can cause this type of failure.An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GROSHONG NXT 5F DUAL-LUMEN REVERSE TAPER PICC BASIC TRAY
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key12049015
Report Number3006260740-2021-02395
Device Sequence Number17042136
Product Code LJS
UDI-Device Identifier00801741035401
UDI-Public(01)00801741035401
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K023374
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number7927505
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2021
Initial Date Received by Manufacturer 06/04/2021
Supplement Date Received by Manufacturer07/12/2021
Initial Report FDA Received Date06/23/2021
Supplement Report FDA Received Date08/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-