| Model Number |
EM100-A |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Report confirmed.The device was tested and the maximum temperature was measured to be 237°f.The likely cause of failure could not be determined.It was also noted that the device was sticking and it stalled.There is no evidence that any other failures during analysis were found that may have contributed to this event.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with no reported malfunction.No patient impact reported.Repair request escalated to product event due to customer indication that this report is a complaint.Additional information received that the device was overheating and there was no patient or staff impact.
|
| |
|