Model Number 71158022
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during an unspecified trauma surgery, a "rusty" appearance was found in the surface of three d-rad smart pack 4h right wide plate ((b)(4)).A surgical delay of less than 30 min occurred due to this event, and the procedure was concluded with a smith and nephew backup right wide plate.No patient injury was reported related to this event.
 
Additional Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The tip of the driver inside the kit has signs of rust.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have reason to suspect that the product failed to meet product specifications at the time of manufacture.This event was evaluated through our internal nonconformance process for corrective action.A health hazard evaluation was also performed.The evaluation determined that, based on the biocompatibility evaluation, the rust/corrosion reported in the complaint is deemed to present low biological risk with the transient tissue contact of this instrument.Therefore, the worst case would be a local adverse tissue reaction not likely to be perceptible to the patient above and beyond the normal local reactions due to the surgery and implantation.The potential probable cause for this event is likely a manufacturing process issue.Based on this investigation, the need for corrective action is indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
D-RAD SMART PACK 4H RIGHT WIDE PLATE
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12052602
Report Number1020279-2021-05378
Device Sequence Number19708746
Product Code HRS
UDI-Device Identifier00885556364598
UDI-Public00885556364598
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K132296
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71158022
Device Catalogue Number71158022
Device Lot Number14EM19889R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2021
Initial Date Received by Manufacturer 06/09/2021
Supplement Date Received by Manufacturer09/02/2021
Initial Report FDA Received Date06/23/2021
Supplement Report FDA Received Date09/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-