MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0545FCS
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2021
Type of Reportable Event Malfunction
Event or Problem Description
Olympus medical systems corp.(omsc) was informed by the physician that during an unspecified procedure using the subject device, the probe was broken off into the patient and an unspecified error occurred.The fragment was retrieved.The intended procedure was completed with another device.There was no patient injury reported.
 
Additional Manufacturer Narrative
The subject device was returned to olympus (b)(4).However, the subject device was not returned to omsc, because it was a known phenomenon.Therefore, we conducted a survey based on the content of the proposal and the attached photo.It was confirmed that the probe was removed from the attached photo to the complaint information by olympus (b)(4).It is possible to infer the cause from the results of past similar investigations, therefore it was determined that an investigation using equipment with similar structure or similar device is unnecessary.The device history record for the serial number indicated no anomaly with the event-related items below.- [inspection] high-frequency output - [inspection] appearance the exact cause of the event couldn¿t be exclusively identified.However based on the past similar case, the probe broken was possibly occurred due to contact with a surgical instrument or the non-insulated area of grasping section.The detailed mechanisms are shown below. during output in seal & cut mode, the probe came in contact with hard tissue, metal or a surgical instrument.Consequently, a scratch was made on the probe.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch and the error occurred.The probe broke when added load. the distal end of the tissue pad was worn away because ¿seal & cut¿ output was activated while grasping nothing (including the case the user kept activating after cutting tissue).The tissue pad was worn away, causing the non-insulated of the grasping section to touch the probe.¿seal & cut¿ output was activated under this situation, then the scratches indicating that the probe and grasping section were in contact with each other were made.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch and the error occurred.The probe broke when added load.The above device handling has warned in the instruction manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP TYPE S
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12055265
Report Number8010047-2021-07909
Device Sequence Number3503204
Product Code GEI
UDI-Device Identifier04953170383526
UDI-Public04953170383526
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K172610
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTB-0545FCS
Device Lot Number0YK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2021
Supplement Date Received by Manufacturer12/27/2021
Initial Report FDA Received Date06/23/2021
Supplement Report FDA Received Date12/27/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-