• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US 9.5X55 POLYAXIAL SCREW; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US 9.5X55 POLYAXIAL SCREW; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 482619555
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device retained by hospital.
 
Event or Problem Description
It was reported that during final tightening, a xia serrato polyaxial screw the head detached from the screw shaft.No adverse consequences or medical intervention were reported.
 
Additional Manufacturer Narrative
Visual, dimensional, functional and material analysis could not be performed as the device was not returned.Device and complaint history records review could not be performed as a valid lot code was not provided and could not be obtained.Per email correspondence with rep, metal fragments may have remained in the patient but no abnormality was found on the post-op x-ray.Torque wrench and anti torque key were used during final tightening.Torque wrench was reported to have fully inserted into the blocker.Rod was fully reduced before final tightening and surgeon did not appear to apply off axis or excessive force during final tightening or struggle during final tightening.It is unknown if the blocker was cross-threaded, if the angle at which surgery was being performed was difficult and if bottom of the tulip contacted bone.Surgical technique advises user to pay extra caution if the rod is not horizontally placed into the screw head, if the rod is high in the screw head or if an acute convex or concave bend is contoured into the rod.Since the device was not returned, an exact cause of the reported event could not be determined.Potential causes include: rod not horizontally placed into the screw head, over angulation on screw, application of excess cantilever force leading to tulip disengagement, etc.
 
Event or Problem Description
It was reported that during final tightening, a xia serrato polyaxial screw the head detached from the screw shaft.No adverse consequences or medical intervention were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9.5X55 POLYAXIAL SCREW
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key12057335
Report Number3005525032-2021-00030
Device Sequence Number19708802
Product Code NKB
UDI-Device Identifier07613327351200
UDI-Public07613327351200
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K170496
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number482619555
Device Catalogue Number482619555
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/26/2021
Supplement Date Received by Manufacturer09/27/2021
Initial Report FDA Received Date06/24/2021
Supplement Report FDA Received Date10/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-