Model Number 66020914
Device Problem False Alarm (1013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/04/2021
Event Type  malfunction  
Event Description
It was reported that, during npwt with renasys go, the 300ml canister is causing the pump to display "full alarm" even though the canister is empty.The canister is exchanged to solve the problem.Patient has not been harmed.
 
Manufacturer Narrative
The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.Inappropriate device alarms and alarms that occur in the absence of a pump malfunction will not directly cause or contribute to serious injury or death, even in a clinical setting.Additional conditions must also exist (e.G.Surgical / treatment conditions; patient factors; the unavailability of back-up or alternate therapies for greater than 24 hours) before an injury could possibly occur, and even under those conditions, the risk of patient harm is low.This event is considered not reportable pursuant to 21 cfr part 803.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS GO CANISTER 300ML
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
MDR Report Key12058124
MDR Text Key258127550
Report Number8043484-2021-01564
Device Sequence Number1
Product Code OMP
UDI-Device Identifier00040565127323
UDI-Public00040565127323
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66020914
Device Catalogue Number66800914
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/04/2021
Initial Date FDA Received06/24/2021
Supplement Dates Manufacturer Received08/24/2021
Supplement Dates FDA Received08/24/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-