• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 250MM BAR; PIN, FIXATION, THREADED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 250MM BAR; PIN, FIXATION, THREADED Back to Search Results
Model Number 71062250
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2021
Event Type  malfunction  
Event Description
It was reported that, during an external fixation, two (2) jet-x 250mm bar's carbon fiber had splintered, and tore the glove of one of the surgeon making the adjustment.It is mentioned that this splinter could have stabbed anyone needed to access this part of the construct, including the patient, resident or attending surgeon, but no one was harmed as consequence of this problem.Fixation was resumed, after a non-significant delay, with a back-up device.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.The pictures provided were reviewed and could not confirm the splintered, but the indention in the bar appears to be from use.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This is a single use device, damage from misuse, excessive cleaning or rough handling are likely probable causes of the reported event.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
250MM BAR
Type of Device
PIN, FIXATION, THREADED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12058157
MDR Text Key258128605
Report Number1020279-2021-05419
Device Sequence Number1
Product Code JDW
UDI-Device Identifier03596010486691
UDI-Public03596010486691
Combination Product (y/n)N
PMA/PMN Number
K994143
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 08/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71062250
Device Catalogue Number71062250
Was Device Available for Evaluation? No
Date Manufacturer Received08/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-