Model Number 71631070
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during trauma surgery, a trigen long hexdriver was noticed to be stripped, outside the patient.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was returned and evaluated.A visual inspection of the returned long hexdriver confirmed the device is dull and stripped at the ends.The device also shows signs of extensive use.A review of complaint history did not reveal any additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG HEXDRIVER
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12058263
Report Number1020279-2021-05421
Device Sequence Number4752236
Product Code HSB
UDI-Device Identifier03596010519269
UDI-Public03596010519269
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71631070
Device Catalogue Number71631070
Device Lot Number19DM25536
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2021
Initial Date Received by Manufacturer 06/10/2021
Supplement Date Received by Manufacturer07/14/2021
Initial Report FDA Received Date06/24/2021
Supplement Report FDA Received Date07/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-