Model Number 16-02-80 |
Device Problem
Electrical Shorting (2926)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to a defective compressor.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova(b)(4) received a report that a heater-cooler system 3t tripped the mains power fuse.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H.10: the root cause is hole in the evaporator due to the stirring flow in the tanks causing water entering the cooling circuit and damaging the compressor.Corrective action is already in place for this issue.The erosion kit that includes the new design of the pump head of the stirrer motor to prevent a water flow directed to a narrow area of the evaporator coil.The issue occurred almost two years after device upgrade.However, most probably the condition of the evaporator was already compromised before the upgrade to such an extent that the copper was rather degraded.
|
|
Event Description
|
See initial report.
|
|