Catalog Number SGC0702
Device Problems Failure to Advance (2524); Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/08/2021
Event Type  malfunction  
Manufacturer Narrative
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the soft tip damage.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with an mr grade of 3-4.Attempts to advance the steerable guiding catheter (sgc), up to the femoral vein were unsuccessful.Resistance was felt with the anatomy.The sgc was removed without issue.The soft tip was slightly damaged, and a kink was noted on the sgc shaft.A new sgc was used.One clip was implanted, reducing mr to 1.There were no adverse patient effects and no clinically significant delay during the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no lot specific product issue.All available information was investigated, and the reported failure to advance was a result of procedural circumstance/operational context.The reported deformation due to compressive stress (soft tip/shaft) was cascading event that resulted due to failure to advance the sgc device.There is no indication of product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12059723
MDR Text Key259411547
Report Number2024168-2021-05383
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/04/2022
Device Catalogue NumberSGC0702
Device Lot Number10204U139
Was Device Available for Evaluation? No
Date Manufacturer Received08/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-