MAUDE Adverse Event Report: ZIMMER DENTAL HEAL COLLAR 3.5X3.5, 3MM; HEALING ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL HEAL COLLAR 3.5X3.5, 3MM; HEALING ABUTMENT Back to Search Results
Model Number HC333
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Patient identifier is not provided / unknown.Patient age is not provided / unknown.Patient weight is not provided / unknown.
 
Event or Problem Description
Customer reported that they believe the healing abutment has damaged threads.They were unable to seat it in the implant.The procedure was completed with a different healing abutment.
 
Event or Problem Description
No additional or corrected information to report.
 
Additional Manufacturer Narrative
This report is being submitted to report additional information.The following sections are being reported: d9: device availability h2: if follow-up, what type h3: device evaluated by manufacturer h6: type of investigation h6: investigation findings h6: investigation conclusions h10: additional narratives/data the reported device was returned and the reported event is confirmed following visual and functional testing.Based on the evaluation, device malfunction has occurred.Additionally, there is no existing nonconformance / capa / hhe/d / ie / product hold against the reported device that did or could cause or contribute to the reported event.Monthly post market trending review identified no adverse trends or actionable trends for the reported event or device.Zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming products.Therefore, based on the available information, the product was within specifications and conforming when it left zimmer biomet.Dhr review was completed for the subject lot number.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review by lot number was performed for similar events no other similar complaint was identified.The reported event could not be recreated due to the nature of the dental device and event and the complaint is related to the functional performance of the device.A definitive root cause could not be identified.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEAL COLLAR 3.5X3.5, 3MM
Common Device Name
HEALING ABUTMENT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key12062893
Report Number0002023141-2021-01631
Device Sequence Number17248754
Product Code NHA
UDI-Device Identifier00889024014978
UDI-Public(01)00889024014978(17)231114(10)2018110793
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K111852
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/14/2023
Device Model NumberHC333
Device Catalogue NumberHC333
Device Lot Number2018110793
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/07/2021
Supplement Date Received by Manufacturer10/27/2021
Initial Report FDA Received Date06/24/2021
Supplement Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-