• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. NS VIS ADPT GUIDE LGNP KIT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. NS VIS ADPT GUIDE LGNP KIT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number V0200109
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, that during a tka surgery, a ns vis adpt guide lgnp kit did not work properly; the tibia fit was insufficient.Surgery was completed, without significant delay, by using a back-up device.The patient was not harmed beyond what has been described.
 
Additional Manufacturer Narrative
H3, h6: the associated device, used in treatment, was not returned for evaluation but the reported event could be confirmed through the engineering evaluation.There was over segmentation in multiple areas of the tibia during the design process.The engineer will be retrained by the segmentation subject matter experts to procedure (b)(4): visionaire tka segmentation standards.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The potential probable cause for this event is a design error.Based on this investigation, the need for corrective action is indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NS VIS ADPT GUIDE LGNP KIT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12063595
Report Number1020279-2021-05435
Device Sequence Number4764764
Product Code JWH
UDI-Device Identifier00885556656563
UDI-Public00885556656563
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K170282
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2021
Device Model NumberV0200109
Device Catalogue NumberV0200109
Device Lot Number00199219V1
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/10/2021
Supplement Date Received by Manufacturer10/13/2021
Initial Report FDA Received Date06/24/2021
Supplement Report FDA Received Date10/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-