MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SURESHOT TARGETER; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SURESHOT TARGETER; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 71692801
Medical Device Problem Code Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during a surgery, a sureshot targeter showed an error message when it was connected to the interface (the targeter was broken).Surgery was resumed, after a non-significant delay, with a back-up device (radiolucent drill).Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible probable cause could include but not limited to software not programmed correctly, software not updated, connection or mechanical issues.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESHOT TARGETER
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12063691
Report Number1020279-2021-05440
Device Sequence Number17334568
Product Code OLO
UDI-Device Identifier03596010648167
UDI-Public03596010648167
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K092497
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71692801
Device Catalogue Number71692801
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/11/2021
Supplement Date Received by Manufacturer08/03/2021
Initial Report FDA Received Date06/24/2021
Supplement Report FDA Received Date08/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-