Model Number 16-02-80
Device Problem Insufficient Cooling (1130)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/25/2021
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to refrigerant leak from the patient side cooling coil and a bubbling sound could be heard.Through follow-up communication livanova learned that the center performed routine disinfection cycle on a weekly basis and this is not in accordance with the instruction for use which prescribes device disinfection every 14 days.The corrosive action of disinfectants may lead to degradation and may resulted in a hole in the cooling coil.The hole in the cooling coil causes refrigerant leak and water entering the refrigeration cycle that damages the cooling compressor.Thus, the increased frequency of disinfection cycles applied by the customer likely exacerbated the above mentioned effect leading to a premature failure of the device.The customer decided to scrap the machine and replacing it with another one.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t could not cool properly during a procedure.The customer swapped out the machine and continued the case with no further issues.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key12064977
MDR Text Key258434666
Report Number9611109-2021-00341
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 06/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/25/2021
Initial Date FDA Received06/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-