• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL SA UNIVERSAL MODULAR ELECTRIC/BATTERY SINGLE TRIGGER HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL SA UNIVERSAL MODULAR ELECTRIC/BATTERY SINGLE TRIGGER HANDPIECE Back to Search Results
Catalog Number 89-8507-400-10
Medical Device Problem Codes Mechanical Problem (1384); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/31/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).The device was not returned to the manufacturer at the date of this report.A follow-up medwatch will be submitted once the investigation is completed or if any additional information is received.
 
Event or Problem Description
It was reported that the universal modular electric/battery single trigger handpiece, part number 89-8507-400-10, serial number (b)(4), was jamming and then got stopped while reaming during a hip replacement.A delay of 30 to 45 minutes in the surgery procedure was reported.The reason for the delay is unknown.There was no additional harm or injury to patient/operator reported.
 
Event or Problem Description
It was reported that the universal modular electric/battery single trigger handpiece, part number 89-8507-400-10, serial number (b)(6), was jamming and then got stopped while reaming during a hip replacement.A delay of 30 to 45 minutes in the surgery procedure was reported.The reason for the delay is unknown.There was no additional harm or injury to patient/operator reported.
 
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Universal modular electric/battery single trigger handpiece, part number 89-8507-400-10, serial number (b)(6) was not returned for complaint investigation.Therefore, the device could not be visually inspected in an effort to confirm the defect.Design history record review was performed and no issue was discovered during the manufacturing process that could explain the defect reported.A follow-up medwatch will be submitted if the product is returned or if additional information is received.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The motor was seized.Additionally, there was a leakage issue at the battery support.The motor, screws, and seals were replaced and resolved the reported issue.The potted wired controller, battery support, and seals were updated.Dhr review was performed.Device was 70 months old and is not out of box failure.Device is used for treatment.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL MODULAR ELECTRIC/BATTERY SINGLE TRIGGER HANDPIECE
Common Device Name
UNIVERSAL MODULAR ELECTRIC/BATTERY SINGLE TRIGGER HANDPIECE
Manufacturer (Section D)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
plan-les-ouates
geneva 1228
SZ  1228
Manufacturer (Section G)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
plan-les-ouates
geneva 1228
SZ   1228
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12065771
Report Number0008031000-2021-00022
Device Sequence Number4000199
Product Code HAB
UDI-Device Identifier00889024354609
UDI-Public(01)00889024354609(11)150302(10)5006009
Combination Product (Y/N)N
Initial Reporter CountryZA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number89-8507-400-10
Device Lot Number5006009
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/07/2021
Supplement Date Received by Manufacturer08/24/2021
02/07/2023
Initial Report FDA Received Date06/25/2021
Supplement Report FDA Received Date08/26/2021
02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/02/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-