• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY CMW - 9610921 VMP + DEPUY CMW 1 50G; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY CMW - 9610921 VMP + DEPUY CMW 1 50G; BONE CEMENT : BONE CEMENT Back to Search Results
Catalog Number 3002050
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint states: ¿bone cement was damaged and physician couldn`t use them at all.No consequences for the patient reported.¿ the device was returned for investigation.The returned device was not in its original complete packaging.The transit cap for the syringe has fractured around 30% of its diameter.This confirms the complaint description.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the cement was damaged and physician couldn`t use them at all.No consequences for the patient reported.
 
Manufacturer Narrative
Product complaint #
=
> (b)(4) investigation summary
=
> the complaint states: ¿bone cement was damaged and physician couldn`t use them at all.No consequences for the patient reported.¿ the device was returned for investigation (see attachment ¿(b)(4) photos.Pdf¿).The returned device was not in its original complete packaging.The transit cap for the syringe has fractured around 30% of its diameter.This confirms the complaint description.The fracture appears spiral and indicates possible overtightening during packing or additional stress during transit.Two new torque machines were installed and validated in 2017 under capa-002454.Em-014315 of capa-002454 confirms that complaints relating to transit caps in the 6 months after the new machines came online reduced by 80%.Dva-104409-fde rev 10 was reviewed, and this possible failure mode is included in lines 98 to 103 (see attachment ¿(b)(4) extract from dva-104409-fde,pdf¿).In each case the risk is considered as low as possible and cannot be further mitigated.There have been 0 other complaints for transit cap issues in the past 12 months.In conclusion this appears to be a random component failure, and it is not possible to determine root cause.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary the number of complaints received for this failure mode will continue to be monitored and product updates/ recommendations will be implemented at the post market surveillance review dependent upon occurrence ratings.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as required.Device history lot
=
> device history reviewed: 0 non-conformances on this lot number.Final micro and sterility tests passed.All qc release specifications met.(b)(4) released.Lot expiry date: 30 november 2021 h10 additional narrative: added: d4 (expiration date).
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> the device associated with this complaint was received for examination.Visual examination of the returned device found the ampule broken, incomplete packaging was received.As the packaging layer containing the liquid cement ampule was not returned it is not possible to determine or establish a root cause.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot
=
> a search of the depuy nonconformance (nc) quality system found two nonconformance recorded on this batch that was related to process controls and did not impact on the overall product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VMP + DEPUY CMW 1 50G
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK  FY4 4QQ
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12066665
MDR Text Key258381663
Report Number1818910-2021-13569
Device Sequence Number1
Product Code LOD
Combination Product (y/n)N
Reporter Country CodePL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 04/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Catalogue Number3002050
Device Lot Number9264682
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/25/2021
Initial Date FDA Received06/25/2021
Supplement Dates Manufacturer Received02/21/2022
06/07/2022
06/23/2022
Supplement Dates FDA Received02/24/2022
06/14/2022
06/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-