• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; PLATE Back to Search Results
Model Number 25-750-14-09
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 05/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Event or Problem Description
It was reported that a plate broke due to patient possibly grinding their teeth and was removed.
 
Additional Manufacturer Narrative
An investigation was performed using a stereo microscope revealed tensile cracks.Further observation determined there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
PLATE
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM  78570
MDR Report Key12067790
Report Number9610905-2021-00056
Device Sequence Number3926470
Product Code MQN
UDI-Device Identifier00888118045492
UDI-Public(01)00888118045492
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K032442
Exemption Number2017029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25-750-14-09
Device Lot Number33140485
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2021
Supplement Date Received by Manufacturer09/17/2021
Initial Report FDA Received Date06/25/2021
Supplement Report FDA Received Date09/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age21 YR
-
-