• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA28-80/I16-40
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Separation Problem (4043)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2 with duraply.
 
Event or Problem Description
The patient was initially implanted with an afx bifurcated stent graft and a non-endologix aortic extension to treat an abdominal aortic aneurysm (aaa).Approximately two (2) years post initial procedure, during a routine follow up, a computed tomography (ct) scan identified a type 3a endoleak with separation between the implanted devices.This initial procedure is outside the indications of use (off-label) due to the use of adjunctive devices not compatible with afx system per the ifu.The physician elected to implant an infrarenal aortic extension to resolve this event.The patient was reported as fine post secondary procedure.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.There are no other equivalent adverse events/incidents for this lot number existing within the endologix complaint handling system.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident could not be completed.No medical records or medical imaging relevant to the adverse event/incident was received by endologix.Due to the absence of relevant medical records and reported medical imaging; device, use, procedure, and/or anatomy relatedness to this adverse event/incident could not be evaluated.During the investigation, endologix found reasonable evidence to suggest the device was implanted outside the indications of use due to the presence of adjunctive devices not compatible with the afx system per device ifu.This is a contributing factor to the reported event.The final patient status was reported as being stable.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: h6 evaluation result codes; remove 3233.H6 evaluation conclusion codes; remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key12068384
Report Number2031527-2021-00246
Device Sequence Number11751989
Product Code MIH
UDI-Device Identifier00818009014467
UDI-Public(01)00818009014467(17)191130
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Other,Recall
Type of Report Initial,Followup
Report Date (Section B) 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2019
Device Model NumberBEA28-80/I16-40
Device Lot Number2015853-005
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/10/2021
Supplement Date Received by Manufacturer06/10/2021
Initial Report FDA Received Date06/25/2021
Supplement Report FDA Received Date07/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0007-2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-