• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 60% stenosed target lesion was located in the mildly tortuous superficial femoral artery.A 5.0 x 60, 75cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 9 atmospheres for 2 seconds, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient was fine.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 60% stenosed target lesion was located in the mildly tortuous superficial femoral artery.A 5.0 x 60, 75cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 9 atmospheres for 2 seconds, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient was fine.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: mustang 5.0 x 60, 75cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 21mm distal of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present on the shaft of the device.No other issues were identified during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12072140
Report Number2134265-2021-08151
Device Sequence Number4987908
Product Code FGE
UDI-Device Identifier08714729793977
UDI-Public08714729793977
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/09/2023
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0026492491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2021
Initial Date Received by Manufacturer 06/11/2021
Supplement Date Received by Manufacturer08/10/2021
Initial Report FDA Received Date06/25/2021
Supplement Report FDA Received Date08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-