| Model Number |
24674 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 60% stenosed target lesion was located in the mildly tortuous superficial femoral artery.A 5.0 x 60, 75cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 9 atmospheres for 2 seconds, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient was fine.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 60% stenosed target lesion was located in the mildly tortuous superficial femoral artery.A 5.0 x 60, 75cm mustang balloon catheter was advanced for dilatation.However, during second inflation at 9 atmospheres for 2 seconds, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: mustang 5.0 x 60, 75cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 21mm distal of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present on the shaft of the device.No other issues were identified during analysis.
|
| |
|
Search Alerts/Recalls
|