Model Number 383401
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The returned pump s/n (b)(4) was received damaged and front case was cracked.The device was manufactured in 2007 and this the first time this serial number was returned for service therefore the reported event was determined to be caused by customer mishandling of the device.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.
 
Event or Problem Description
The customer reported that the device had fire damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEY,PUMP W/CLMP,REFB
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
MDR Report Key12072815
Report Number1282497-2021-10384
Device Sequence Number4166224
Product Code LZH
UDI-Device Identifier10884521144019
UDI-Public10884521144019
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number383401
Device Catalogue Number383401
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/17/2019
Initial Report FDA Received Date06/26/2021
Patient Sequence Number1
-
-