| Model Number |
382400 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump s/n: (b)(4) was received and could not be tested to confirm the reported issue due to physical damage.During visual evaluation it was noted that all internal components were damaged.The damaged parts were replaced.The pump was tested and passed all functional test.The service history record was reviewed.The device was manufactured in 2013.This is the first time this device has been returned for service.The damaged components were determined to be caused by customer mishandling of the device.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device had high feed rate.
|
| |
|