Model Number 72203981
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that the healicoil anchor thread peeled on the cortex at initial insertion after following surgical insertion technique during the shoulder arthroscopy.The standard healicoil punch was used to prepare the insertion site (3.8mm).Anchor was removed.The procedure was completed with a smith and nephew backup device.There was no significant delay.No other complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A review of the material found there are requirements for conformance and a certificate of material analysis is required.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALICOIL PK 4.5MM W/ ULTRA TAPE BLUE
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key12073139
Report Number1219602-2021-01426
Device Sequence Number3975904
Product Code MBI
UDI-Device Identifier00885554032123
UDI-Public00885554032123
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K113294
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2025
Device Model Number72203981
Device Catalogue Number72203981
Device Lot Number2061637
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/07/2021
Supplement Date Received by Manufacturer07/28/2021
Initial Report FDA Received Date06/27/2021
Supplement Report FDA Received Date07/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-