| Model Number |
7210551 |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the t-max ii was not holding in place at full 90.It is unknown whether the event happened during surgery and if there was patient involvement.It is unknown if there was a delay or if a backup device was available and how the issue was resolved.No other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product.A cart was returned with the device.The left strut is heavily leaking fluid from the top.A functional evaluation was performed.The device will not hold weight.The left strut failed.The reported failure was confirmed and the root cause has been associated with a seal failure.A review of device records shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
| |
|