Upon checking the device on-site, the dispatched dräger service engineer could confirm the observed shutdown of automatic ventilation.The engineer identified problems with the ventilator motor and replaced it, consequently.The device passed all consecutive tests and could be returned to use.Evaluation of the returned original motor performed in the manufacturer's lab revealed that wear-and-tear related abrasion of the collector disc had resulted in development of positions where the motor does not provide mechanic power due to contact interrupts to the carbon brushes; speed fluctuations will be the consequence.These speed fluctuations may result in a deviation between intended and expected piston position and can cause an application of false tidal volumes and/or damages to the ventilator unit.To prevent from that the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.Dräger finally concludes that the device behaved as specified upon the malfunction of a single component after almost 14 years of use; no patient consequences have been reported.The repair exchange will fully solve the device problem.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|