ARTHREX, INC. SUPERMX COMPRESSION STAPLE, 18W X 18L; STAPLE, FIXATION, BONE Back to Search Results
Model Number SUPERMX COMPRESSION STAPLE, 18W X 18L
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during arthroscopy surgery the clip did break off intraoperative.The broken off piece has been retrieved from the patient.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.No further information received.
 
Additional Manufacturer Narrative
Complaint confirmed.Two unpackaged ar-8719mx01818 were received for investigation.Visual inspection identified that both staples had broken at one end, with a single leg returned measuring approximately 0.61".The other fragment was not returned for analysis.Both staple bodies were assessed using digital micrometer id: 224 and were found to meet design specifications.One of the two staple bodies presented surface damage/chipping across its top width.The cause of the event remains undetermined.Information as to the method of bone preparation, as well as the bone quality encountered, was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERMX COMPRESSION STAPLE, 18W X 18L
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key12076060
Report Number1220246-2021-03330
Device Sequence Number14364339
Product Code JDR
UDI-Device Identifier00888867300941
UDI-Public00888867300941
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K203180
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUPERMX COMPRESSION STAPLE, 18W X 18L
Device Catalogue NumberAR-8719MX-1818
Device Lot Number009912816
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/08/2021
Supplement Date Received by Manufacturer06/08/2021
Initial Report FDA Received Date06/28/2021
Supplement Report FDA Received Date10/01/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-