| Model Number |
LOW-PROFILE AC IMPLANT DEVICE |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the tightening suture of the device broke on the collarbone side during an acromioclavicular repair surgery.The broken piece was retrieved from the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Visual evaluation showed that the fiberlink ends are frayed and tigerlink has returned in three pieces, all ends are frayed.Also, the top hat button was not returned.The cause is undetermined, however a likely cause is user-applied excessive forces.
|
| |
|