• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. LOW-PROFILE AC IMPLANT DEVICE; WASHER, BOLT NUT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. LOW-PROFILE AC IMPLANT DEVICE; WASHER, BOLT NUT Back to Search Results
Model Number LOW-PROFILE AC IMPLANT DEVICE
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that the tightening suture of the device broke on the collarbone side during an acromioclavicular repair surgery.The broken piece was retrieved from the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Complaint confirmed.Visual evaluation showed that the fiberlink ends are frayed and tigerlink has returned in three pieces, all ends are frayed.Also, the top hat button was not returned.The cause is undetermined, however a likely cause is user-applied excessive forces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOW-PROFILE AC IMPLANT DEVICE
Common Device Name
WASHER, BOLT NUT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key12076084
Report Number1220246-2021-03331
Device Sequence Number19759104
Product Code HTN
UDI-Device Identifier00888867224247
UDI-Public00888867224247
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K052776
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberLOW-PROFILE AC IMPLANT DEVICE
Device Catalogue NumberAR-2370
Device Lot Number11039542
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/10/2021
Supplement Date Received by Manufacturer06/10/2021
Initial Report FDA Received Date06/28/2021
Supplement Report FDA Received Date09/22/2021
Patient Sequence Number1
-
-