| Model Number |
IBAL TKA,TIB BEARING TRL SZ 1, 11MM |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the ar-623-t112 and ar-623-t111, broke during use.The case was completed without incident by using a new device.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed, evaluation shows the device broke in two.The bridge between the two halves broke off and the fragment was not returned.The inferior side of the device is also dented.A likely cause of the event is user-applied mechanical forces.
|
| |
|