• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3660
Device Problems Battery Problem (2885); Wireless Communication Problem (3283)
Patient Problems Failure of Implant (1924); Inadequate Pain Relief (2388)
Event Date 06/10/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
It was reported that the patient's ipg became depleted and stopped providing therapy.It is unknown if the ipg depleted normally or prematurely.As a result, surgical intervention may take place at a later date to address the issue.
 
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
 
Event Description
Additional information received stated that the patient's ipg stopped communicating with external devices following an unrelated surgery wherein surgery mode was not enabled.A manufacturer representative met with the patient and confirmed the ipg was inoperable.
 
Manufacturer Narrative
A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
 
Manufacturer Narrative
Patient weight added to the report.
 
Event Description
Additional information received stated that the patient underwent surgical intervention wherein the ipg was explanted and replaced.Post-operatively, stimulation therapy was restored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key12076554
MDR Text Key258675080
Report Number3006705815-2021-03091
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067031419
UDI-Public05415067031419
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Remedial Action Other
Type of Report Initial,Followup,Followup,Followup
Report Date 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/13/2022
Device Model Number3660
Device Catalogue Number3660
Device Lot NumberA000102188
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number1627487/06/02/2017/001-C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight61
-
-