Model Number 0502859030
Medical Device Problem Codes Poor Quality Image (1408); Optical Problem (3001)
Health Effect - Clinical Codes Menorrhagia (4508); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was blurry image.
 
Event or Problem Description
It was reported that there was blurry image.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.The technical service report is attached (see communication log), and indicates: replacement of damaged optics type probable root cause: ¿ incorrectly assembled optical train ¿ damage to optical train ¿ end of life wear-out ¿ shipping damage ¿ use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD 10MM 30A LAPAROSCOPE, AC
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key12076668
Report Number0002936485-2021-00340
Device Sequence Number4169587
Product Code GCJ
UDI-Device Identifier07613327060171
UDI-Public07613327060171
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K910132
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502859030
Device Catalogue Number0502859030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2021
Initial Date Received by Manufacturer 06/02/2021
Supplement Date Received by Manufacturer06/02/2021
Initial Report FDA Received Date06/28/2021
Supplement Report FDA Received Date08/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-