MAUDE Adverse Event Report: DEPUY MITEK LLC US VAPR3 GENERATOR; ELECTROSURGICAL SYSTEM GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VAPR3 GENERATOR; ELECTROSURGICAL SYSTEM GENERATOR Back to Search Results
Model Number 225021
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The lot number was unknown.Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the device had internal error.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: error code - 200, physical damage, loose components - connector/coupler, minor scratches on the unit.The parts were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event or Problem Description
It was reported by the affiliate in that preoperatively to a shoulder arthroscopy procedure on (b)(6) 2021, it was observed that the vapr3 generator device had an unspecified internal error.During in-house engineering evaluation, it was determined that the device had an error code - 200.Another like device was used to complete the procedure without delay.There were no adverse patient consequences reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR3 GENERATOR
Common Device Name
ELECTROSURGICAL SYSTEM GENERATOR
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key12076966
Report Number1221934-2021-01987
Device Sequence Number17250561
Product Code GEI
UDI-Device Identifier10886705009107
UDI-Public10886705009107
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K113545
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number225021
Device Catalogue Number225021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2021
Initial Report FDA Received Date06/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-