• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GENTHERM MEDICAL, LLC HEMOTHERM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GENTHERM MEDICAL, LLC HEMOTHERM Back to Search Results
Model Number 400CE
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint (b)(4) received.After speaking with biomed, it was concluded that the issue was with the cool return solenoid.Perfusionist did not want to return the device to use until biomed could change the solenoid.Replacement part was sent to the customer.There were no allegations of patient harm.To date, no parts have returned for evaluation.
 
Event or Problem Description
Customer states that this hemotherm is continually having issues with the cool reservoir filling while heating and the heat reservoir emptying during procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOTHERM
Common Device Name
HEMOTHERM
Manufacturer (Section D)
GENTHERM MEDICAL, LLC
12011 mosteller rd
cincinnati OH 45241
Manufacturer Contact
angie gegner
12011 mosteller rd
cincinnati, OH 45241
MDR Report Key12079577
Report Number1516825-2021-00006
Device Sequence Number11759506
Product Code DWC
UDI-Device Identifier10613031860221
UDI-Public(01)10613031860221
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K122813
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number400CE
Device Catalogue Number86022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2021
Supplement Date Received by Manufacturer06/03/2021
Initial Report FDA Received Date06/28/2021
Supplement Report FDA Received Date12/23/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/20/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-