| Model Number |
1550300-23 |
| Medical Device Problem Codes |
Inflation Problem (1310); Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562); Device Damaged by Another Device (2915); Difficult to Advance (2920); Device Dislodged or Dislocated (2923); Activation Failure (3270)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/10/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties and treatment appear to be related to circumstances of the procedure as it is likely the device interacted with the anatomy and previously implanted stent during advancement causing the reported difficulty to advance and device damaged another (previously expanded stent mangled).An attempt was made to inflate the balloon; however, the reported interaction with the anatomy and/or previously implanted stent likely caused reported material rupture and subsequent inflation problems with activation failure.The device was inflated to nominal pressure expanding the distal portion of the stent prior to the reported rupture.The partial expansion and reported interaction with the mangled previously expanded stent likely caused the reported stent dislodgement and foreign body in patient.During removal resistance was felt likely due to the inability of the ruptured balloon to fully deflate prior to removal causing the reported difficulty to remove and subsequent material separation.The separation occurred outside of the anatomy and the device was removed without issue.An attempt was made to advance a balloon catheter; however, they were unable to re-engage and the artery was left untreated.There is no indication of a product quality issue with respect to manufacture, design or labeling.Na.
|
| |
|
Event or Problem Description
|
|
It was reported the procedure was to treat a de novo lesion in the mid circumflex coronary artery.When advancing the 3.0x23 mm xience sierra stent delivery system (sds), resistance was felt with the anatomy and the stent got caught on a previously implanted stent, possibly mangling the previously implanted stent.An attempt was made to inflate the balloon; however, the balloon did not inflate properly.The balloon only inflated on the distal end.It is suspected the proximal end of the balloon leaked.The stent did not deploy and dislodged inside the previously implanted stent.Resistance was felt during removal of the sds.Once the sds was outside of the anatomy, the proximal shaft separated.An attempt was made to get back in the vessel with a balloon to try to expand the stent struts inside the previously implanted stent; however, the artery could not be re-engaged, therefore the lesion was left untreated.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|