• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941816350
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 3.50 x 16mm synergy xd drug-eluting stent was selected for use.However, during the preparation prior to insertion, the shaft broke.The procedure was completed with another of same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy xd mr us 3.50 x 16mm, stent delivery system was returned for analysis.Examination of the crimped stent via scope found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was measured and the result was within max crimped stent profile measurement the balloon cones were reviewed via scope, and no issues were noted.The balloon wings were tightly wrapped and evenly folded.The balloon had not been subjected to positive pressure.A visual and tactile examination of the hypotube found a break in the hypotube at 41.5cm distal to the distal end of the strain relief.Multiple kinks were identified along both sections of the hypotube.Examination of the tip via scope found no damage to the tip.The shaft polymer extrusion including the distal and mid-shaft sections were examined via scope and tactile examination.There was no evidence of damage.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that shaft break occurred.A 3.50 x 16mm synergy xd drug-eluting stent was selected for use.However, during the preparation prior to insertion, the shaft broke.The procedure was completed with another of same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12081708
Report Number2134265-2021-08233
Device Sequence Number3348791
Product Code NIQ
UDI-Device Identifier08714729981046
UDI-Public08714729981046
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/10/2023
Device Model NumberH7493941816350
Device Lot Number0026867282
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2021
Initial Date Received by Manufacturer 06/14/2021
Supplement Date Received by Manufacturer07/22/2021
Initial Report FDA Received Date06/28/2021
Supplement Report FDA Received Date07/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-