| Model Number |
H7493941816350 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.50 x 16mm synergy xd drug-eluting stent was selected for use.However, during the preparation prior to insertion, the shaft broke.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy xd mr us 3.50 x 16mm, stent delivery system was returned for analysis.Examination of the crimped stent via scope found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was measured and the result was within max crimped stent profile measurement the balloon cones were reviewed via scope, and no issues were noted.The balloon wings were tightly wrapped and evenly folded.The balloon had not been subjected to positive pressure.A visual and tactile examination of the hypotube found a break in the hypotube at 41.5cm distal to the distal end of the strain relief.Multiple kinks were identified along both sections of the hypotube.Examination of the tip via scope found no damage to the tip.The shaft polymer extrusion including the distal and mid-shaft sections were examined via scope and tactile examination.There was no evidence of damage.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.50 x 16mm synergy xd drug-eluting stent was selected for use.However, during the preparation prior to insertion, the shaft broke.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|