|
It was reported that two t7 linear driver shaft w/ ao qc was broken, bent.It is unknown if the event happened during a surgical procedure, if there was a surgical delay, backup device available or patient involvement.
|
|
H6: the device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned t7 linear driver shaft w/ ao qc confirmed the device is fractured on the end rendering the device inoperable.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch for the same failure mode.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|